Specification of EUTECH Waterproof Portable TN100
Alat Laboratorium Umum
EUTECH Waterproof Portable TN100
Specifications
Product Specification
Description
Principle of Operation Nephelometric non-ratio (conformance to
ISO 7027 and DIN EN 27027)
Turbidity Range 0 to 1000 NTU
Resolution 0.01 NTU (0 to 19.9 NTU)
0.1 NTU (20.0 to 99.9 NTU)
1 NTU (100 to 1000 NTU)
Accuracy
± 2% of reading ± 1 LSD for 0 to 500 NTU
± 3% of reading ± 1 LSD for 501 to1000 NTU
Repeatability ± 1% of reading or ± 0.01 NTU
No. of Calibration Points 4 points (automatic)
Response Time < 6 seconds for full step change
Calibration Standards 0.02 NTU; 20.0 NTU; 100 NTU; 800 NTU
Standardization EPA-approved polymer-based primary standards
Light Source Life > 1,000,000 tests
Light Source Infrared-emitting diode (850nm wavelength)
Special Functions Auto-Ranging; Self-diagnostic
Power Requirements 4 x 1.4V batteries, > 200 hours
Order Information
Order No.
Item
ECTN100IR Waterproof TN 100 Infra-Red Turbidimeter (NTU) with calibration kit (ECTN100CALKT), 3 sample vials
(ECTN100CUVKT) and carrying case
01X621001 TN100CALKIT Calibration Kit Set for TN100IR (0.02, 20.0, 100, 800 NTU), 4 x 60 mL