Technical Specifications
method & detection UV oxidation (partial) & conductivity
sample volume flow approx.14 mL/min
period of measurement 2 sec (online mode) | 1 min (offline mode)
measurement range 0.5 to 1000 ppb
detection accuracy ± 0.5 ppb or 1 % (whatever is greater)
ports 1 for sample, 1 for SST (basic) | 1 for sample, 3 for SST (pro)
related norms USP <643>, EP 2.2.44, DIN EN 1484
ambient parameters relative humidity: 10 to 95 %, temperature: 10 to 45 °C
interface ports analog, according to Namur NE43
dimensions, weight, power 300 x 200 x 500 mm, 14 kg, 110 - 230 V, 50 Hz, 60 Watt
Technical Specifications
Medium Requirements
type of sample medium pure to ultrapure water
medium conductivity < 2 µS/cm (upgradable to < 10 µS/cm)
inlet pressure < 1 bar
sample temperature < 50 °C