sql

ProDIGITAL ODO/CT Cable Assembly YSI

Rp. 100
Last Updated
16 Mar 2023
Country Origin
United States
Minimum Order
1 Unit
Category
YSI

Specification of ProDIGITAL ODO/CT Cable Assembly YSI

Accuracy

Temperature - ±0.2 °C

Dissolved Oxygen - 0 to 200%: ±1% of reading or 1% saturation, whichever is greater 200 to 500%: ±8% of reading 0 to 20 mg/L: ±0.1 mg/L or 1% of reading, whichever is greater 20 to 50 mg/L: ±8% of reading

Conductivity - 0 - 100 mS/cm: ±0.5% of reading or .001 mS/cm, whichever is greater 100 - 200 mS/cm: ±1.0% of reading

Specific Conductance - 0 - 100 mS/cm: ±0.5% of reading or .001 mS/cm, whichever is greater 100 - 200 mS/cm: ±1.0% of reading User selectable reference temperature (15 to 25 °C; default 25 °C) and compensation coefficient (0 to 4%/°C; default 1.91%)

Salinity - ±1.0% of reading or ±0.1 ppt, whichever is greater

Total Dissolved Solids (TDS) - Calculated from specific conductance and a user-selectable TDS multiplier (0.30 to 1.00; default 0.65)

Seawater Density - Sigma is calculated from salinity, temperature, and pressure (depth) Sigma T is calculated from salinity and temperature

Resistivity - ±0.1% Full Scale

Equipment used withProDSS, ProSolo
Measurement Range

Temperature - -5 to 70 °C (temperature compensation range for DO mg/L measurement: -5 to 50 °C )

Dissolved Oxygen - 0 to 500%, 0 to 50 mg/L

Conductivity - 0 to 200 mS/cm

Specific Conductance - 0 to 200 mS/cm

Salinity - 0 to 70 ppt

Total Dissolved Solids (TDS) - 0 to 100 g/L

Seawater Density - 0.0 to 50.0 sigma, sigma T

Resistivity - 0 to 2 Mohms

Resolution

Temperature - 0.1 °C or 0.1 °F (user selectable)

Dissolved Oxygen - 0.01 mg/L and 0.1%, or 0.1 mg/L and 1% (user selectable)

Conductivity - 0.001, 0.01 or 0.1 mS/cm

Specific Conductance - 0.001, 0.01, 0.1 mS/cm

Salinity - 0.01 ppt

Total Dissolved Solids (TDS) - 0.001, 0.01, 0.1 g/L

Seawater Density - 0.1 sigma or sigma T

Resistivity - 0.001, 0.01, 0.1

Size

0.97 in. width x 10.29 in. length (2.46 cm x 26.14 cm)

Unit of MeasureTemperature - °C, °F, K; DO - % saturation, % saturation local, mg/L, ppm; Conductivity - µS/cm, mS/cm; Specific Conductance - µS/cm or mS/cm; Salinity - ppt or PSU; Total Dissolved Solids (TDS) - mg/L, g/L, kg/L; Seawater Density - Sigma, Sigma T; Resistivity - ohm-cm, kohm-cm, Mohm-cm
Warranty

2-year probe and cable assembly; 2-year extended warranty ODO sensor caps; 1-year ODO sensor caps