Technical Specifications
method & detection UV oxidation & NDIR
(enhanced oxidation; configurable with reagent)
sample volume 20 mL loop (low range), 5 mL loop (high range)
period of measurement 6 to 15 min
measurement range 2 to 5000 ppb (low range), 15 ppb to 20000 ppb (high range)
detection accuracy ± 3 ppb or 2 % (whatever is greater)
ports 2 for samples | 1 for reagent
related norms USP <643>, EP 2.2.44, DIN EN 1484, 21 CFR part 11, IP 45 DIN 32645,
ambient parameters relative humidity: 10 to 95 %, temperature: 10 to 45 °C
interface ports analog (4 - 20 mA)
dimensions, weight, power 400 x x 600 mm, 24 kg, 110 - 230 V, 50 Hz, 70 to 105 Watt 300
Technical Specifications
Medium Requirements
type of sample medium from ground water to ultra pure water
sample temperature 10 to 50 °C
inlet pressure < 1.5 bar
compressed gas air, oxygen, nitrogen | free of carbon dioxide
inlet pressure of gas nitrogen: 1 to 1.5 bar | oxygen/air: 3.5 to 5 bar
Alat Laboratorium Umum
Harga UniTOG membraPure:CALL